Programmable/Down Scalable Digital Pressure Transmitter Type P1217

Programmable/Down Scalable Digital Pressure Transmitter Type P1217

The wide application field of pressure transducers is guaranteed by the high accuracy and the rugged design.

The stainless steel membrane is completely vacuumsealed, extremely burst resistant and applicable for all standard media for hydraulics, pneumatics, environmental engineering, process technology, semiconductor technology and automotive engineering. Thus the use for standard applications of mobile hydraulics and other application areas is covered.

Due to the manufacturing process all pressure transducers are being individually pressure- and temperaturetested.
The production occurs due to the requirements of DIN EN ISO 9001:2008.

APPLICATIONS:

  • Pneumatics / Hydraulics
  • Mechanical Engineering
  • Vehicle – Technology
  • Mobile – Hydraulics
  • Water treatment
  • Food- and Beverage industry
  • General industrial applications
  • Pharmaceutical Industry

Title 1

Title 2

Title 3

Title 4

Title 5

Technical data

Standard
pressure ranges * ) ( bar)
-1_____ 0,1_____ 0,3_____ 0,6_____ 1_____ 1,6_____ 2,5_____ 4_____ 6_____ 10_____ 16_____ 25_____ 40_____ 60_____ 100 _____ 160_____ 250_____ 400_____ 600_____ 1000_____ 1600_____ 2000
Overload ( bar) *) Max. 1.5 times / 1.2 times – depending on pressure range
Burst pressure ( bar) *) 2 times / 1.5 times – depending on pressure range
Kind of pressure Vacuum, gauge pressure, absolute pressure on request
Wetted parts Stainless steel
Weight ( g) Approx.. 200 g
Supply voltage ( VDC ) 10…30 V at 4..20 mA / 14…30 V at 0…10V
Output signals and max. load 4…20mA, 2 wire Ra<=(Ub-10V)/20mA 0...5 V, 3 wire Ra>5 ohm
0…10V, 3 wire Ra>10 ohm

 

Digital optional
Others on request

Adjustability of zero

 

 

Straightforward zero correction by using a magnet or via interface and
PC programming kit
Adjustability of span 1:4 with pressure ranges (FS) via interface and software
Adjustability time constant via interface and software
Sample rate ˃250 Hz
Response time (10 … 90 %) ≤ 10 ms
Accuracy **) ≤ 0,35 % BFSL
(Including non-linearity, zero point and full scale error, hysteresis, non-linearity and repeatability)
Compensation measurement and adjustment for vertical mounting position
Non-linearity *** ) % FS ≤ 0,25
Repeatability % FS ≤ 0,1
Long-term stability % FS ≤ 0,1 1-year stability at reference conditions
Permissible temperatures  
Media temperature  -20….+ 80 ( 120) °C
Ambient temperature Ambient temperature -20….+ 80°C
Storage temperature Storage temperature -20….+ 80°C
Compensated temp. range -20…+ 80 °C
Temperature coefficient zero % FS ≤ 0,1 / 10K
Temperature coefficient FS % FS ≤ 0,1 / 10K
Coefficient FS % FS ≤ 0,1 / 10K
CE-conformity
 
Pressure equipment directive 97/23/EG
EMC directive 2004/1 08 EG
Shock resistance g 100 to IEC 60068-2-27 mechanical
Vibration resistance g 20 to IEC 60068-2-6 resonance
Wiring protection  
Over voltage 32 VDC
Short-circuit strength Out+ / UB- (for 1s)
Reverse polarity UB+ / UB-

More info

The zero can be set easily with a magnet within ± 10 % of the nominal range.

To correct the zero point, hold a permanent magnet– a pin board magnet, for example – at the position marked on the pressure transmitter (i.e. a letter in a circle) for ½ to 2 ½ minutes after the power has been switched on. To correct the zero, atmospheric pressure is applied. Offsets for previously set values for initial and ultimate pressures will be corrected automatically by the device. A magnetic field applied outside of this set again.

PARAMETERIZATION

Parameterization of measuring range (1:4), zoom, span and adjustment of characteristic diagram are possible with service tool or PC adapter and software. Programmable devices are only available with Electrical connection by Binder M12 (Binder series S763) or by cable.
The service tool or the PCF software are available accessory.

SERVICE

The sensors are free of maintenance.

NOTE

We compiled this operating instruction carefully. Nevertheless, it was not possible to take all possibilities of application into account. If this data sheet should lack the solution of your special task, please don’t hesitate to contact us.

SAFETY INFORMATION

During installation, putting into service and operation of the pressure sensors, it is necessary to observe the relevant safety regulations that are in force in the country of the user (as for example, DIN VDE 0100). Errors excepted; subject to alterations in the sense of technical improvement.


Standard in stock

Following ranges are availble from Tempress stock: 0-1/0-1,6/0-2,5/0-4/0-6/0-10/0-16/0-25/0-40/0-60/0-100/0-160 Bar.

The stock items are all with media contacted parts in stainless steel, G1/4B process connection and electrical connection MVS/Form A DIN EN 175301-803.

Other connectors on request.

Programmable/Down Scalable Digital Pressure Transmitter Type P1217

CONSTRUCTION:

The compensation and adjustment is carried out electronically. Thus the pressure transmitters have a very low total error and a very good long-term stability. The measuring cell is characterised by its high long-term resistance and long-term stability.
With the precision of modern electronics, the measured data can be captured accurately. Even the programming of the pressure transducers by the user can be realised on a service tool or PC programming kit. The graduation of the measuring range and the zero point can be set up through the digital interface. Furthermore sensor data can be readout from the device. By using permanent magnets the adjustment of the zero point can easily and securely be done at any time.

SPECIAL FEATURES:

Accuracy

  • < 0.50% FS
  • < 0.35% BFSL

Robust construction

  • Wetted parts of stainless steel
  • Protection category IP 65

Programmable by PC programming kit or service tool

  • Zero point (offset)
  • Down scalable 4:1
  • Zoom and SPAN
  • Characteristics / Output options
    (inverse, square rooted, special forms)

Straightforward zero correction by using a magnet

Output signals

  • 4 … 20 mA
  • 0 … 10 V
  • 0 … 5V
  • Digital

Process connections

  • Standard to chart
  • Others on request

Diaphragm seals available

  • According to specific datasheets

Figure similar depending on model


Connections Diagram

2-wire current (4…20mA)*

3- wire voltage (0…10V , 0…5V)*

M12x1 (Binder series S763)

2 wire : 1 + , 2; 3 wire : 1+, 2 Gnd , 3+ out/ signal*

Cable : 2 wire : red + , black; 3 wire : red + , black -, white out,*

* ) Others on request

Standard articles

Click on the article to automatically insert into the contact form below

To view our standard articles please visit from a desktop

No items found.

Single contact

(1 contact point)

Avaliable contact types:

KN,KM,KE and KI

Contact Function : Break

1

Contact Function : Close

2

Contact Function : Shift

3

Contact Function Separate Circuits : Break

4

Contact Function Separate Circuits : Close

5

Double contact

(2 contact points)

Avaliable contact types:

KN,KM,KE and KI

Shift contact

(2 contact points switch function)

Avaliable contact types:

KN and KM

Double contact

(2 contact pointers seperate Circuits)

Avaliable contact types:

KN and KM

Triple contact

(3 contact pointers)

Avaliable contact types:

KN and KM

Triple contact

(3 contact pointers)

Avaliable contact types:

KN and KM

Do you need assistance?

Vi står altid til rådighed til at hjælpe dig, uanset hvilke spørgsmål du måtte have. Tøv ikke med at kontakte os gennem kontaktformularen, eller find din kontaktperson her

Thank you for reaching out. We'll be in touch as soon as possible.
Oops! Something went wrong while submitting the form.