Inductive Contact System Type KI

Inductive Contact System Type KI

INDUCTIVE CONTACT – TYPE KI

  • Inductive contact system for reliable contact function for pressure gauges and thermometers
  • Vibration proof
  • For dry and liquid filled instruments
  • 1 or 2 contact points
  • Break or close function
  • Working voltage 5-25 VDC
  • 100 mA load (transistor outlet)
  • Long durability

Title 1

Title 2

Title 3

Title 4

Title 5

Specifications

Contact system Type KI, inductive
Instrument size: DN 100 and DN 160 on request
Contact type KI

 

Number
Functions

Inductive

 

1 or 2 functions
Break or close

Voltage:
Own consumption:

 

Max. load:

5-25 VDC
≥3 mA – activated
≤1 mA – non-activated
≤100 mA
Min. measuring range: Pressure gauge:
1 bar (15 psi)
Thermometer:
80°C (180°F)
Working range
Minimum Working range
10-90% of Span
5% of Span
Accuracy ±2% at span ≤ 4 bar
±1,5% at span ≥ 4 bar
Hysteresis As specific instrument class
Ambient temp. -20+70°C
Electrical
connection
Plug, ISO 4400 (DIN43650-A)
2 plugs by 2 contacts
Load Depends on the relays used

Inductive Contact System Type KI

Application

The inductive contact type functions via a proximity type of switching, i.e. without mechanical contact.

Due to the proximity type of switching, the switching accuracy and the high service life, inductive contacts may be used for almost all industrial applications, and should be given special preference for areas under a risk of an occasional explosive atmosphere.

If suitable isolating amplifiers are used, the operating material complies with the “i” type of protection, i.e. intrinsically safe which is classified under EEx ib II C T6 and permitted in areas which are under risk of explosion.

Inductive contacts always require the use of a relay or similar.


Standard articles

Click on the article to automatically insert into the contact form below

To view our standard articles please visit from a desktop

No items found.

Single contact

(1 contact point)

Avaliable contact types:

KN,KM,KE and KI

Contact Function : Break

1

Contact Function : Close

2

Contact Function : Shift

3

Contact Function Separate Circuits : Break

4

Contact Function Separate Circuits : Close

5

Double contact

(2 contact points)

Avaliable contact types:

KN,KM,KE and KI

Shift contact

(2 contact points switch function)

Avaliable contact types:

KN and KM

Double contact

(2 contact pointers seperate Circuits)

Avaliable contact types:

KN and KM

Triple contact

(3 contact pointers)

Avaliable contact types:

KN and KM

Triple contact

(3 contact pointers)

Avaliable contact types:

KN and KM